CMM Verification: ISO 10360, ASME B89, and Keeping Your CMM in Tolerance
A coordinate measuring machine is often the final authority in your inspection process. When the CMM says a part is good, it ships. That authority depends on the machine performing within its specification, and the only way to know is to verify it.
CMM verification tests the machine’s length measurement and probing performance against calibrated artifacts and documents the result. This guide covers CMM technology, the governing standards, and the performance issues that verification and interim checks catch.
CMM Technology Overview
CMM Types and Configurations
- Bridge CMMs. The most common configuration. A moving bridge spans the table and carries the probe. Good accuracy across a wide range of part sizes.
- Cantilever CMMs. Open on three sides for easy loading of small parts.
- Horizontal arm CMMs. Used for large, heavy parts such as car bodies and large castings.
- Gantry CMMs. Built for very large parts, with the structure supported independently of the floor.
Probe Systems and Measurement Principles
Touch-trigger probes record a point each time the stylus contacts the part. Scanning probes maintain contact and collect continuous data, which improves form measurement. Optical and laser sensors collect points without contact. Each probe type has its own error characteristics, and each stylus configuration must be qualified.
Software and Algorithm Considerations
The CMM measures points; software turns those points into features. Fitting algorithms, such as least squares or minimum zone, can produce different results from the same data. Most modern CMMs also apply software error compensation, a map of the machine’s geometric errors. If that map no longer matches the machine because of a crash, a move, or wear, every result shifts.
Verification Standards
ASME B89.4.1 and ISO 10360
ISO 10360 is the international series for CMM acceptance and reverification. ISO 10360-2 covers length measurement error, and ISO 10360-5 covers probing error.
Many US contracts still reference ASME B89.4.1, which evaluates performance including volumetric tests with a ball bar. Newer US specifications typically reference ASME B89.4.10360.2, which aligns with ISO 10360-2. Check which standard your customer or quality system requires before scheduling verification.
Volumetric Accuracy Testing
A three-axis CMM has 21 geometric error sources: six per axis (position, straightness in two directions, and three rotations) plus three squareness errors between axes. Volumetric testing evaluates their combined effect.
Under ISO 10360-2, a calibrated length artifact, such as a step gauge or gauge block set, is measured at five lengths in seven positions within the measuring volume, three times each. The 105 results are compared to the machine’s maximum permissible error, usually expressed in the form A + L/K micrometers.
Probe Qualification Procedures
Probe qualification is a routine user task, separate from verification. Each stylus and probe angle is qualified against a calibrated reference sphere to establish its effective diameter and position. Qualification is repeated whenever a stylus is changed, after any collision, and at the start of measurement sessions as your procedures require. The reference sphere needs its own calibration.
Performance Issues
Thermal Drift and Environmental Effects
Temperature is the largest source of CMM error. Gradients across the machine, airflow from HVAC vents, sunlight, and heat from nearby equipment distort the structure and shift results. Many CMMs specify an environmental range for their stated accuracy; outside that range, the specification does not apply.
Probe Wear and Damage
Stylus tips wear and flatten with scanning, especially on abrasive materials. Collisions bend styli and damage probe heads. A damaged probe may still qualify on the sphere while measuring incorrectly on real features.
Machine Geometry Errors
Crashes, relocation, settling foundations, and bearing wear change the machine’s geometry. The compensation map no longer matches, and errors appear that the software cannot correct until the machine is re-verified and, if needed, recompensated.
EML’s CMM Services
EML Calibration is ISO/IEC 17025:2017 accredited by Perry Johnson Laboratory Accreditation. We perform coordinate measuring machine verification with NIST-traceable artifacts, along with calibration of the gauge blocks, reference standards, and inspection tools that support your CMM program.
- Artifact-based testing. Verification uses calibrated length artifacts measured within your machine’s working volume.
- Performance documentation. Results are compared to the machine’s specification, with as-found data and stated measurement uncertainty.
- On site, where the machine lives. CMMs are verified in place. Our mobile labs serve facilities within 250 miles of Nashville.
Frequently Asked Questions
What is the difference between CMM verification and CMM calibration?
Verification tests whether the machine performs within its specified maximum permissible error, using calibrated artifacts. Recompensation, which updates the machine’s error map, is a separate service usually performed when verification shows the machine is out of specification.
How often should a CMM be verified?
Annual verification is common, with interim checks between verifications on an interval set by your quality system. Verify again after any collision, relocation, or major repair.
What is ISO 10360-2?
ISO 10360-2 is the international standard for testing a CMM’s length measurement error. It specifies measuring a calibrated length artifact at five lengths in seven positions, three times each, and comparing the results to the machine’s maximum permissible error.
Schedule Your CMM Verification
Send us your CMM make, model, measuring volume, and the standard your customer requires. Request a Quote